MedBAC ™ Specifications

Home > ScienceHospital compressed air MedBACMedBAC ™ Specifications

MedBAC ™

for stationary use after compressed air preparation

Online monitoring of the compressed air after the treatment plant

 

Medium: compressed air

 

Components and measuring ranges:

  • Oil content 0.001 ... 9.999 mg / m³ (droplets, aerosols, vapors)
  • Oxygen 0,0 ... 25,0 Vol.%
  • Water vapor 0.5 ... 23700 ppm (v / v)
  • Carbon dioxide 0 ... 1000 ppm (v / v)
  • Carbon monoxide 0 ... 300 ppm (v / v)
  • Nitrous gases 0.0 ... 5.0 ppm (v / v)
  • Sulfur dioxide 0.0 ... 5.0 ppm (v / v)
  • Standard basis: DIN EN ISO 7396-1: 2
  • Operating temperature: 0 to + 50 ° C
  • Operating pressure: max. 16 bar

evaluation:

  • Measuring unit equipped with evaluation electronics
  • TFT screen display with graphical data display and event memory built into front panel
  • Degree of protection: IP54
  • Signal outputs: normally closed
  • Power supply: 115 - 230V AC, 50/60 Hz
  • Housing: impact-resistant plastic
  • Operating temperature: 0 to + 50 ° C

For measurement in accordance with ISO 8573: 2001, an inlet section of 10 x DN riser, a measuring arc with probe connection and outlet section of 3 x DN is required.

The MedBAC calibration is performed by comparison with reference standards or reference standard devices that have been standardized in an accredited calibration center and are thus attributed to the national standards with which the Physikalisch Technische Bundesanstalt (PTB) represents physical units in accordance with the international standard system (SI) ,